Current Output - ABB MagMaster Instruction Manual

Electromagnetic flowmeter
Hide thumbs Also See for MagMaster:
Table of Contents

Advertisement

3.4.5 Current Output – Fig. 3.20 and 3.21
Information.
• Output is fully programmable – see Programming Guide.
• Output is electrically separated from all other MagMaster connections.
• External isolators are not normally required and may significantly limit accuracy if used.
Information. For Multidrop HART installations, remove 'HART Link' and connect HART
systems directly to IC2: this allows the analog output function to be retained.
Receiver No. 2 – Reverse
Receiver No. 1 – Forward
Fig. 3.20 Current Output Connections: Standard
Reverse +ve
Fig. 3.21 Current Output Connections: Dual Current Option
3 ELECTRICAL INSTALLATION...
MagMaster Transmitter
HART link
+ve
Common
IC+ IC–
IC2 IC–
Alternative
Normal
HART connections
where applicable
I n f o r m a t i o n . M u l t i d r o p
HART mode cannot be used
with this configuration.
+ HART
MagMaster Transmitter
Forward +ve
Common
IC–
IC2
HART connections
(where applicable)
IC2
IC+
IC –
IC2
IC+
IC –
15