5.F. Cleaning the Injectors
We recommend thoroughly cleaning the injectors every 30 days.
1.
Use a prepared solution of 70% ethanol to flush the injector(s) three times. Prepare at least 10ml of ethanol
for each injector to be flushed.
2.
Place the container of 70% ethanol into the bottle holder. The inlet tube should be inserted so that it reaches
the bottom of the container.
3.
Ensure the Waste Collection Tray is in position and empty.
4.
From the Tools screen (SettingsTools; Tools icon at top right of screen), select "Injectors". Tap the Prime
button to initiate priming.
5.
Allow the solution to sit in the fluid path for 30 minutes before flushing the injectors.
6.
From the Tools screen (SettingsTools; Tools icon in upper right corner of the title bar), select "Injectors".
Tap the Flush button to show the Flush window.
7.
Select 9 flush cycles and press Start Flush to wash the lines with 70% ethanol.
8.
Empty and clean the Waste Collection Tray, and place it back into position.
9.
Place a container of deionized water on the bottle holder with the inlet tube inserted to reach the bottom of
the container. Prepare at least 10ml of deionized water for each injector to be flushed.
10.
From the Tools screen (SettingsTools; Tools icon in upper right corner of the title bar), select "Injectors".
Tap the Flush button to initiate flushing.
11.
Select 9 flush cycles, and press Start Flush to wash the lines with deionized water.
12.
Empty and clean the Waste Collection Tray, and place it back into position.
13.
Flush the injector 3 times with air. A small volume of water will remain in the injector after the air purge.
5.G. Cleaning the Waste Collection Tray
The maximum volume capacity of the Waste Collection Tray is approximately 50ml. Carefully remove the Waste
Collection Tray from its position to avoid liquid spills.
After each prime and flush sequence, discard any fluid from the Waste Collection Tray, and rinse it with deionized
water.
5.H. Removing or Replacing Inlet and Outlet Tubing
Replacement tubing assembly is available. See Section 7.E for ordering information.
1.
Turn OFF the instrument. Disconnect the AC Power Adaptor.
2.
Push the Microplate Sample Tray to the back of the instrument and push the Optical Head from the Home
position toward the center of the instrument (Figure 54).