5.D. Cleaning the Optical Head
1.
Turn OFF the instrument. Disconnect the AC Power Adaptor. Do not attempt to clean the Optical Head
while the instrument is ON.
2.
Push the Microplate Sample Tray to the back of the instrument and push the Optical Head from the Home
position toward the center of the instrument (Figure 54).
Fi
gure 54. Interior of the GloMax
3.
Loosen the hand screws and remove the Injector Tip Holder from the Optical Head. Facing the front of the
instrument, push upward and then leftward (as viewed from the bottom of the Tip Holder to release the
Tip Holder from the Optical Head (Figure 55). Completely remove the Injector Tip Holder to take out the
Aperture.
Figure 55. Remove the Injector Tip Holder.
4.
Place your hand underneath the Optical Head. Grasp the Aperture and pull it toward you (Figure 56).
Note: Do not touch the bottom surface of the Optical Head.
40
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
TM470 · 4/16
Navigator instrument.
®
Optical head
Microplate sample
tray and cover
www.promega.com