position with the START key and
initiate therapy by pressing the
START key again.
5.6.2 Isolated internal rota-
tion/external rotation
1. Complete the mechanical settings
and adjust the carriage to the patient
as described in section 4.2.
2. Press the MENU key on the pro-
gramming unit to access level 1 (M1).
3. Press the parameter key for adduc-
tion
or abduction
press the +/- keys to set the carriage
to the desired position to be main-
tained during the isolated rotation
movements.
4. Next set the desired range of motion
for internal/external rotation by first
pressing the parameter key for
internal rotation
values with the +/-keys.
5. Set the external rotation value in the
same way
.
Note!
- For an exclusive rotation protocol,
motor A for adduction/abduction
must be turned off as described in
steps 6 to 8 below.
- Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
6. Press the MENU key on the pro-
gramming unit to access level 3 (M3).
7. Press the Motor A ON/OFF
parameter key to activate the param-
eter.
8. Deactivate motor A by pressing the
parameter key Motor A ON/OFF
again or the "–" (minus) key. The
check mark in the circle next to the
symbol must have disappeared.
9. Save the settings with the STOP key,
set the carriage to the starting
position with the START key and
initiate therapy by pressing the
START key again.
1. First adjust an angle of 90° for
and
Note!
- For an exclusive elevation protocol,
- Having programmed the ranges of
and set the
2. Complete the mechanical settings
71
5.6.3 Isolated elevation
(flexion)
external rotation and deactivate
motor B as follows:
a. Press the MENU key on the
programming unit to access
level 1 (M1).
b. Press the parameter key for
external rotation
the carriage to the rotation posi-
tion of 90°, using the + / - keys.
motor B for the rotation movement
must be turned off as described in
steps c to e below.
motion, you can program additional
options, such as pause or speed.
c. Press the MENU key on the
programming unit to access
level 3 (M3).
d. Press the Motor B ON/OFF
parameter key to activate the
parameter.
e. Deactivate motor B by pressing
the parameter key Motor B
ON/OFF again or the "–" (minus)
key. The check mark in the circle
next to the symbol must have
disappeared.
f. Save the entered values by press
the STOP key.
and adjust the carriage to the patient
as follows:
a. Adjust an angle of 90° for horizon-
tal adduction/abduction
indexing knob (12).
b. Secure and hold the motion
element at the holding tube for
length adjustment and open the
locking screw (14) for height
Œ
adjustment
. Select a height at
which the pivot points of motor A
and move
, using