Follow these steps to program pauses:
• Access menu level 2, then press the
parameter key to select the special
function
.
• The display indicates the symbol for
the special function and the selected
pause durations.
top: abduction/external rotation
pause
bottom: adduction/internal rotation
pause (see display image)
The pause duration for
abduction/external rotation appears
in a dotted frame.
• To change the abduction/external
rotation pause, press the "+" or "-"
key. Select a duration of 10 seconds,
for instance.
• Press the parameter key again
when done. The dotted frame
automatically moves to the line
below, indicating that the adduc-
tion/internal rotation pause can be
adjusted.
The parameter key is only used to
toggle between these two functions.
• To modify the value, press the "+"
(plus) or "-" (minus) key. Select a
duration of 5 seconds, for instance.
• Save the programmed values by
pressing the STOP key, then press
the START key to begin the treat-
ment.
I Therapy timer
Default setting is continuous
operation.
A clock symbol in the upper right-
hand corner of the display identifies
the continuous mode of opera-
tion.The clock indicates the elapsed
therapy time.
In the continuous mode, the device
must be stopped with the STOP key.
However, the therapy timer can be
set in steps of 1 minute to any
value between 1 and 300 minutes.
At the end of the therapy time, the
device automatically stops in the
starting position (maximum adduction
position / halfway between internal
and external rotation).
In this case, a circle replaces the
clock symbol. The circle fills as the
therapy time progresses.
I Speed
The speed can be adjusted between
1 % and 100 % in steps of 1 %.
100 % is equivalent to 230°/minute
Default: 100 %
I New patient
With this function, the CPM device
will move to the home position.
Select the function and press the
START key:
– The carriage moves to the home
position.
– The programmed treatment
parameters are deleted.
– All values stored on the chip card
are deleted.
– The device stops halfway between
the abduction/adduction and
internal/external rotation position.
The "New patient" function (home
position) selects the following
settings:
62