Ormed Artromot S3 Operating Instructions Manual page 64

Motor-operated continuous passive motion device providing motion to the shoulder joint
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 44
LEVEL 4:
I Transport setting
With this function, the carriage will
move to a position optimally suited
for packing the CPM device. Select
the function and press the START
key. The carriage moves to the
transport position. In the top left-
hand corner of the display you will
see
(also refer to section 6.3
Transport).
I Synchronized/
non-synchronized mode
Motors A and B can run in synchro-
nized or non-synchronized mode.
Synchronized:
Motors A and B perform a synchro-
nized movement according to the
shoulder joint's physiological motion
pattern:
Starting from the initial position
(maximum adduction / halfway
between internal and external rota-
tion position), the carriage will first
move to the maximum internal
rotation position. The carriage will
then simultaneously move to the
maximum abduction position and the
maximum external rotation position
and subsequently to the maximum
adduction position and the maximum
internal rotation position. After
reaching this position, the cycle
restarts: maximum abduction and
maximum external rotation.
During operation the synchronized
mode is indicated in the top left-hand
corner by the symbol
Non-synchronized:
Both motors run independently of
each other within the selected ranges
of motion.
To enable the non-synchronized
mode, disable the synchronized
mode.
During operation the non-synchro-
nized mode is indicated in the top
left-hand corner by the symbol
Default: synchronized mode
enabled
Patient hazard 
The synchronized operation is recom-
mended as a general rule. However,
medical and/or therapeutic indications
may call for the non-synchronized
mode. In the non-synchronized mode,
particular caution and attention on the
part of the physician/therapist is
required to rule out any risk for the
patient.
I Total therapy time
Under menu item "total therapy time"
you can view each patient's total
therapy time (duration of all the
patient's therapy sessions).
To delete the stored therapy time:
Press and hold the parameter key
for 5 seconds or select the "New
patient" function.
I Service menu
For service purposes only, refer to
Service Manual.
Reminder:
You save the selected parameter values
by pressing the STOP key.
.
64
Caution!

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents