Ormed Artromot S3 Operating Instructions Manual page 67

Motor-operated continuous passive motion device providing motion to the shoulder joint
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 44
Follow these steps to program the
special function:
• Access menu level 4, then press the
parameter key to select the special
function. The display will indicate:
– the symbol representing the
special function
– the status of the special function
(circle with or without check mark)
– two boxes with the current
percentage of the stop positions
for motor A and B
– a check mark in a circle
• You activate the function with the "+"
key and deactivate it with the "-" key.
In this case it is not possible to
activate/deactive the function by
pressing the parameter key again. A
check mark in the circle identifies the
activated function.
• Next press the parameter key of the
isolation protocol. The dotted frame
moves to the upper percentage value,
allowing you to enter the stop position
for motor A.
• To modify the value, press the "+"
(plus) or "-" (minus) key. Select 75%,
for instance.
• Press the parameter key again when
done. The dotted frame automatically
moves to the lower value, allowing
you to adjust the stop position of
motor B.
• To modify the value, press the "+"
(plus) or "-" (minus) key. Select 25%,
for instance.
• Save the programmed values by
pressing the STOP key, then press
the START key to begin the treat-
ment.
This is what happens:
Abduction/adduction is treated ten
times first (motor A). Motor B is posi-
tioned at 25% of the programmed
range of motion for internal and external
rotation.
Subsequently, internal / external rota-
tion is treated ten times (motor B).
Motor A is positioned at 75% of the
programmed range of motion for
adduction and abduction.
Note!
The percentage values can be
modified only when the special
function is enabled (check mark).
I Therapy documentation adduc-
tion/abduction
This is a special function of
ARTROMOT
®
-S3 Comfort which
allows the entire therapy documenta-
tion to be represented. The carriage
run times as well as the range of
motion of the sessions are recorded.
The collected data are presented
graphically in the form of a coordinate
system (X-axis = range of motion /
Y-axis = time) where the upper curve
illustrates the trend of the abduction
movement and the lower curve the
trend of the adduction movement.
I Therapy documentation internal /
external rotation
This is a special function of
ARTROMOT
®
-S3 Comfort which
allows the entire therapy documenta-
tion to be represented.
The carriage run times as well as the
range of motion of the sessions are
recorded.
The collected data are presented
graphically in the form of a coordi-
nate system (X-axis = range of
motion / Y-axis = time) where the
upper curve illustrates the trend of
the external rotation movement and
the lower curve the trend of the
internal rotation movement.