3. Electrical Connections

Sensor Selection

Select the sensor with a range closest to your process minimum and
maximum values.
• CDCE-90-001 (0.01 cell): 0.055 to 100 µS (10 kΩ to 18 MΩ)
Use the CDCE-90-001 for all resistivity requirements
CDCE-90-01 (0.1 cell):
CDCE-90-1 (1.0 cell):
CDCE-90-10 (10.0 cell):
CDCE-90-20 (20.0 cell):
Wiring Tips:
Do not route sensor cable in conduit containing AC power
wiring. Electrical noise may interfere with sensor signal.
Routing sensor cable in grounded metal conduit will help
prevent electrical noise and mechanical damage.
Seal cable entry points to prevent moisture damage.
Only one wire should be inserted into a terminal. Splice
double wires outside the terminal.
Sensor cable may not be extended over 100 ft.
Caution: Failure to fully open terminal jaws before removing wire may permanently damage instrument.
Wiring Procedure
1. Remove 0.5 - 0.625 in. (13-16 mm) of insulation from wire end.
2. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
3. Insert exposed (non-insulated) wire end in terminal hole until it bottoms out.
4. Release orange terminal lever to secure wire in place. Gently pull on each wire to ensure a good connection.
Wiring Removal Procedure
1. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
2. When fully open, remove wire from terminal.

3.1 Sensor Connections

page 4
1 to 1000 µS
10 to 10,000 µS
100 to 200,000 µS
200 to 400,000 µS
CDTX-90-1 Terminals
Shield (Sensor Gnd)
10
Black (Iso Gnd)
9
White (Temp In)
8
Red (Signal In)
7
CDTX-90-2 Terminals
Shield (Sensor Gnd)
14
Black (Iso Gnd)
13
White (Temp In)
12
Red (Signal In)
11
CDTX-90-3 Terminals
Shield (Sensor Gnd)
14
Black (Iso Gnd)
13
White (Temp In)
12
Red (Signal In)
11
Ultra-
CDCE-90 Series Sensor Ranges
Pure
Pure
Rinse
CDCE-90-20
CDCE-90-10
CDCE-90-1
CDCE-90-01
CDCE-90-001
0.055
1
10
100 200
1,000
(10 kΩ)
(18 MΩ)
Conductivity Range (µS)
CDTX-90
Conductivity/Resistivity
≤ 100 ft. (30 m)
62.50 uS/cm
25.0°C
ENTER
CDCE-90 Series Standard and Certified Cells
10,000
200,000
400,000
2
1