Setting Of The Current Limit; Electronic Fuse; Tracking Function - Hameg HM7044 Manual

Hide thumbs Also See for HM7044:
Table of Contents

Advertisement

S e t t i n g t h e p a r a m e t e r s

3. Setting of the current limit

Current limiting means that a maximum current can be set.
This is e.g. useful in order to protect a sensitive test circuit. In
case of a short in the test circuit the current will be limited to
the value set which will (in most cases) prevent damage.
U
out
U
const
Adjustment of current
The picture shows that the output voltage V
while the current I
increases until the current limit selec-
out
ted I
will be reached. At this moment the instrument will
max
change from constant voltage regulation to constant current
regu-lation. Any further load increase will cause the current
I
to remain stable while the voltage U
out
to zero.
If the current limit set is reached the red LIMIT LED
and indicates that the current regulation is active.
The current limit of the channels I – IV can be set with the
respective rotary control
keypad
:
If the current limit of a channel shall be adjusted with the re-
spective rotary control, the A-LED
mode is activated by operating the A-button
rotary control clockwise the current limit of this channel is
increased and a counterclockwise rotation decreases it. The
faster the rotary control is turned, the higher the increase of
the value of one revolution will be.
If the numeric keypad
is used for parameter setting, the
changes of the current limit will be carried out for all selected
channels. Thereby several channels can be set at the same
time. After selcting the channels to be changed (the respective
CH.-LEDs
are blinking), the required current limit can be
entered. The last entered fi gure can be deleted by operating the
DEL-button
. The entry is fi nished by pushing the A-button
. If no channel is selected (all CH.-LEDs
channels will be set to the entered value.
Entered value
3
100
0.3
1.5
3.0
3000
Examples:
Setting up the current limit of all channels to 2 A:
Push buttons 2
and A
Setting up the current limit of channels I and III to 0.1 A:
Push button SEL
Push buttons I and III
Enter 0.1
and push button A
22
Subject to change without notice
I
I
max
out
remains stable,
out
decreases ultimately
out
glows
as well as with the numeric
has to glow. This setting
. By turning the
do not blink), all
Accepted value
3 mA
100 mA
300 mA
1.5 A
3.0 A
3.0 A

Electronic fuse

In order to provide a still better protection than current limiting
the HM7044 features an electronic fuse.
By operating the FUSE-button
reset. With the selection mode the fuses of several channels
can be activated. If the fuse of a channel is set, this channel
will be disabled immediately as soon as the current limit is
reached. The respective FUSE LED
activated for this channel.

Tracking function

By means of the tracking function several channels can be
linked together. The output voltage as well as the electronic
fuse of the channels I – IV can be tracked.
Pushing the TRACK.-button
The displays of the channels I – IV show respectively a fi gure
between 1 to 4 instead of the nominal values of the ouput volta-
ge and the current limit. The fi rst row indicates the coupling of
the voltage setting and the second row indicates the coupling
of the electronic fuses of channels I – IV. The tracking mode
is left by pushing the TRACK.-button
The tracking function is active not until the same fi gure is
assigned to two or more channels. The fi gures can only be
changed with the respective rotary control
function is active the TRACK.-LED
the tracking function all fi gures have to be set back to their
original values. This is done automatically by operating the
DEL-button
in the tracking mode.
Tracking - Voltage
For the coupling of the output voltage setting the fi rst row of
the displays is relevant. The tracking function for the voltage
setting mode is acitve if at least the fi gures of two channels
were set to the same value. If the voltage of one of these chan-
nels is changed by the rotary control, the voltage of the linked
channels are changed automatically, too.
Example:
Tracking of channel I and II and additionally channel III and
IV:
Push the TRACK.-button
. Adjust one of the following com-
binations by the rotary control
channel I
channel II
channel III channel IV
1
1
2
1
1
3
2
2
1
2
2
3
Push the TRACK.-button
.
Chance the voltages with the rotary controls
Tracking - electronic fuse
If overcurrent occurs at one channel and if the electronic fuse
of this channel is activated, all channels are disabled, which
are linked with this one.
If the channels I, II and III are linked together, for example, and
the electronic fuse is activated for channel I and II, an over-
current (e.g. short circuit) at channel III has no consequences
for the channels linked with this one. But an overcurrent at
channel I or II will disable the channels I, II and III.
the fuse(s) can be set and
glows if the fuse is
the tracking mode is accessed.
again.
. If the tracking
glows. For deactivating
.
2
3
1
3 etc.
.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents