Remington iLIGHT Pro Use And Care Manual page 23

Hide thumbs Also See for iLIGHT Pro:
Table of Contents

Advertisement

Available languages

Available languages

Troubleshooting
PROBLEM
POSSIBLE
CAUSE
The outer ENERGY
The Remington
LEVEL SELECTION
iLIGHT device
DISPLAY lights are
isn't working
flashing alternately and
properly.
the Remington iLIGHT
device is beeping
every ½ second.
The second and
The SKIN
fourth ENERGY
CONTACT
LEVEL SELECTION
SENSORS may
DISPLAY lights are
be stuck.
flashing and the
Remington iLIGHT
device is beeping.
WHAT YOU SHOULD DO
Contact Customer Service for help.
See page 48.
Lift the HAND PIECE from the BASE UNIT and
check the SKIN CONTACT SENSORS. The SKIN
CONTACT SENSORS should slightly protrude
from the sides of the HAND PIECE window.
You should be able to press the SKIN CONTACT
SENSORS down so that they are flush with the
HAND PIECE FLASH WINDOW and they should
spring back to about 2mm when released.
If the SKIN CONTACT SENSORS do not depress
or spring back, they may be stuck and may
require cleaning. See "After Your Remington
iLIGHT Hair Removal System Treatment" in
the TOC for instructions on how to clean your
device. If you still can't get it working, contact
Customer Service for help. See page 48.
44
PROBLEM
POSSIBLE CAUSE
All the lights on
The TREATMENT CAP
the system are
is removed or not
flashing and the
properly attached.
unit is beeping
constantly.
I used the SKIN
Your skin is too dark.
TONE SENSOR
See Figure 1.
and it showed an
YELLOW light. I
You didn't use the
can't unlock the
SKIN TONE SENSOR
system.
correctly.
The SKIN TONE
SENSOR isn't working.
WHAT YOU SHOULD DO
Check to ensure that the TREATMENT
CAP is properly attached to the HAND
PIECE. See Figure 5 for instructions
on how to check that the TREATMENT
CAP is properly attached. When the
TREATMENT CAP is properly attached,
you should hear a "click" and the
lights and audible tones will stop.
You cannot use the system with your
skin tone. Please contact Customer
Service for help. See page 48.
Try using the SKIN TONE SENSOR again
– refer to steps 2b-2f in Figure 3 for
instructions on using the SKIN TONE
SENSOR.
Please contact Customer Service for
help. See page 48.
45

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ilight pro plus quartz

Table of Contents