General Specifications - Fluke 12E User Manual

Table of Contents

Advertisement

General Specifications

Maximum Voltage between any Terminal
and Earth Ground:
 Fuse Protection for mA μA inputs:
 Fuse Protection for A input:
Display:
Temperature:
Operating Altitude:
Temperature Coefficient:
Electromagnetic Compatibility:
Rated Transient Overvoltage:
Relative Humidity:
Relative Humidity, 40 MΩ Range:
Battery Type:
Battery Life:
Size (H x W x L):
Weight:
Certifications:
Safety Compliance:
Overvoltage installation categories per IEC 61010-1, 2000:
CAT I
From high-voltage low-energy sources, e.g., electronic circuits or a copy machine.
CAT II
From equipment supplied from the fixed installation, e.g., TVs, PCs, portable tools and household appliances.
From equipment in fixed equipment installations, e.g., installation panels, feeders and short branch circuits, and lighting
CAT III
systems in large buildings.
1000V
500 mA, 1000 V FAST, Min. Interrupt Rating 20,000 A
10 A, 1000 V FAST, Min. Interrupt Rating 20,000 A
Digital: 4000 count updates 3/sec
Operating: 0 °C to 40 °C,
Storage: -30 °C to 60 °C indefinitely (to -40 °C for 100 hours)
0 to 2000 meters
0.1 X (specified accuracy)/ °C (<18 °C or >28 °C)
Complies with FCC Part 15, Class B, IEC 61326, 3 V/m, performance criterion B
4 kV (1.2 x 50 μs) Peak for measurement Categories I, II, and III.
Noncondensing < 10 °C
90% from 10 °C to 30 °C; 75% from 30 °C to 40 °C
80% from 10 °C to 30 °C; 70% from 30 °C to 40 °C
2 X AA, NEDA 15A, IEC LR6
Alkaline: 500 Hours
180 mm x 89 mm x 51.5 mm (with holster)
425 grams
CMC, 
IEC 61010-1, 2000 CAT I 1000 V, CAT II 600 V, and CAT III 300 V overvoltage standards
The Meter is designed to protect against transients in these catagories:
Multimeter
General Specifications
11

Advertisement

Table of Contents
loading

Table of Contents