Preventive Maintenance—
Provider

Introduction

Properly maintaining the Companion 5 will ensure longer life and
higher performance. Preventative Maintenance is required every 2
years at a minimum.
CAUTION
The Companion 5 contains electrostatic sensitive
components. Do not open or handle except at a
static free workstation. Do not remove cover with-
out electrostatic discharge (ESD) protection.

Maintenance Check List

Perform the following maintenance procedures at least every two
years or more often, as needed. The frequency of the periodic
maintenance should be based on the environment in which the
Companion 5 is used.
• Inspect AC Power cord and plug for damage
• Read and record hour meter
• Check flow rate, concentration, and alarm functions
• Replace air intake filter and HEPA filter

Maintenance Schedule

Preventative Maintenance Check list
Perform the following maintenance procedures at least once
every two years or more often, as needed. The frequency of the
periodic maintenance should be based on the environment in
which the Companion 5 is used. Harsh environments may require
preventative maintenance more frequently than every 2 years.
• Replace air intake filter
• Replace HEPA filter

Oxygen Concentration Check

CAIRE recommends the oxygen concentration be checked peri-
odically as required by the provider. For non-OCSI units, CAIRE
recommends a 6 month interval between concentration checks.
Oxygen Concentrator
PN 14940837 B — 20