Guidance and manufacturer's declaration - electromagnetic emissions
The Model CP910E is intended for use in the electromagnetic environment specified below.
The customer or user of the Model CP910E should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11/EN 55011
RF emissions
CISPR 11/EN 55011
Harmonic emissions
IEC/EN 61000-3-2
Voltage fluctuations /
flicker emissions
IEC/EN 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The Model CP910E is intended for use in the electromagnetic environment specified below.
The customer or user of the Model CP910E should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
Electrical fast
transient/burst
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC/EN 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC/EN 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
Group 1
Class B
Class A
Complies
IEC/EN 60601
Compliance
test level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1 kV for input
±1 kV for input
/output lines
/output lines
±1 kV
±1 kV
differential
differential
mode
mode
±2 kV common
±2 kV common
mode
mode
< 5% U
< 5% U
T
(> 95% dip in U
)
(> 95% dip in U
T
for 0.5 cycle
for 0.5 cycle
40% U
40% U
T
(60% dip in U
)
(60% dip in U
T
for 5 cycles
for 5 cycles
70% U
70% U
T
(30% dip in U
)
(30% dip in U
T
for 25 cycles
for 25 cycles
< 5% U
< 5% U
T
(> 95% dip in U
)
(> 95% dip in U
T
for 5 sec.
for 5 sec.
3 A/m
3 A/m
Electromagnetic environment - guidance
The Model CP910E uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause interference in nearby electronic
equipment.
The Model CP910E is suitable for use in all
establishments, including domestic establishments and
those directly connected to public low-voltage power
supply network that supplies buildings used for
domestic purpose.
Electromagnetic environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
T
typical commercial or hospital
)
T
environment. If the user of the Model
CP910E requires continued operation
T
during power mains interruptions, it is
)
T
recommended that the Model CP910E be
powered from an uninterruptible power
supply or a battery.
T
)
T
T
)
T
Power frequency magnetic fields should
be at levels characteristic of a typical
commercial or hospital environment.