Troubleshooting; Fault - Cause - Remedy; Correct Disposal - DeVilbiss UltraNeb User Manual

Hide thumbs Also See for UltraNeb:
Table of Contents

Advertisement

Available languages

Available languages

6. Troubleshooting

26
Fault
Cause
Red LED
Nebulizer chamber not fitted
comes on
correctly
Nebulizer chamber empty
Leak in nebulizer chamber
Faulty transducer
Electronic fault
Not enough
Nebulizer fluid does not
nebuli zation
match scale
when set to
maximum
Resistance of filter too high
output
Condensation in patient hose
Leak in hoses or lid of
nebulizer chamber
No nebulization
Mains switch at "0" (off)
Not connected to mains power
Faulty mains cable
Mains fuse blown
No nebulizer fluid in
nebulizer chamber
Electronic fault
Plug for heated hose
not connected
Heated hose
Faulty hose
does not
Electronic fault
heat up

Correct disposal

This device may not be disposed of at municipal collection points for waste electrical and
electronic equipment. Please contact DeVilbiss or your local authorised dealer for advice on
disposal of this device in the proper manner. This device is subject to EU Directive
2002/96/EC (WEEE).
Remedy
Insert nebulizer chamber correctly
Be careful that it is properly
positioned.
Refill with nebulizer fluid
Screw transducer firmly into
nebulizer chamber
Replace transducer
Call Sunrise Medical Service
Refill with nebulizer fluid or
remove some
Replace filter
Empty condensation. Reroute
patient hose so that condensate can
flow back into nebulizer chamber
Fit lid tightly and secure with clamp
Connect hoses firmly to nozzles
Turn mains switch to "I" (on)
Connect to mains
Replace mains cable
Replace mains fuse,
use 2 x IEC 127 T 800 L 250 V type
Replenish nebulizer fluid
Call Sunrise Medical Service
Connect plug
Replace hose
Call Sunrise Medical Service

Advertisement

Table of Contents
loading

Table of Contents