Troubleshooting - DeVilbiss EasyFit SilkGel Instructions For Use Manual

Face mask
Table of Contents

Advertisement

7

Troubleshooting

Problem
Pain in face due
to pressure.
Dry eyes.
Therapy pressure
is not reached.
Cause
Mask is too tight.
Mask is too loose.
Mask does not fit.
Mask not adjusted
correctly.
Mask cushion
damaged.
Tube system
leaking.
Air escaping at
pressure
measurement ports.
®
EasyFit
SilkGel Full Face Mask
Remedy
Loosen headgear slightly.
Modify position of forehead support.
Tighten headgear slightly. Modify
position of forehead support.
Use a different mask size. Contact
your equipment provider.
Readjust mask (see "4.3 Adjust
mask" page 18).
Replace mask cushion.
Check push-fit connector and fit of
tube sleeves.
Seal pressure measurement ports
with cap (included).