2. Getting started

1
2
3
4
5
6
7
8
Aerosol therapy must
only be carried out as
directed by a doctor !
20

Connecting to electrical power supply

Note - Make sure that the mains voltage on the unit supplied corresponds with the operating
voltage in your location. If in doubt please contact an electrical specialist or your
local DeVilbiss authorised dealer. If OK, remove the safety label covering the Equip-
ment power socket.
Connect the mains plug to the equipment power socket and to an earthed mains socket
outlet.

Use with nebulizer standard hose

1. Inspect hose to check for tears or damage before using
2. Make sure that the tubing is properly fixed to the UltraNeb connectors.
3. In case of damage the tubing must be replaced. Please contact DeVilbiss or your
authorised dealer
4. The tube should only be disconnected at the cuffs. Do not pull hard or twist the tube as
the tube may become damaged.

Use with heated hose

1. Use only the supplied hose !
2. Malfunctions may arise if other manufacturers heated hoses are used !
3. Connect plug on heated hose to socket at back of nebulizer. The mark on the plug should
be pointing upwards.
4. Put heated silicone hose (instead of 1200 mm hose) to the outlet of the nebulizer cham-
ber lid, continuous filling system or medication cup.
The hose heating switches on automatically when ultrasonic nebulizer starts.
The operating temperature of max. 37° C is reached after about 2 minutes.

Check the operation

After cleaning, before use set the main switch located on the rear panel to "I" (ON).
Unit will start running
Check the operation controls
- nebulizer
- airflow
- timer
Yellow LED shows which level of function is selected. Push +/- to in/decrease function in steps
1. It takes about 2 minutes for heated hose to warm up.
2. To test the alarms, Remove nebulizer chamber from UltraNeb
3. Red LED comes on and alarm sounds
4. The alarm will sound for 12 sec and then fall silent for 4 sec. This will continue for a
period of 2 min before stopping, the Red LED will remain on
5. Replace the nebulizer chamber back into UltraNeb .
6. Push the switch ON/OFF (O/I) the Red LED with go out.
7. The UltraNeb is ready for use again. The green LED is lit indicating stand-by mode.