Cleaning Procedure; Replacement Of Transducer; Replacement Of Bacterial Filter; Cleaning The Single / Multiple-Use Hoses - DeVilbiss UltraNeb User Manual

Hide thumbs Also See for UltraNeb:
Table of Contents

Advertisement

Available languages

Available languages

24

Cleaning procedure

Parts
Nebulizer chamber lid
Continuous filling
equipment
Nebulizer chamber
without transducer
Hose 300 mm
Hose 1200 mm
Heated silicone hose
Bacterial filter
Medication cup
Medication cup lid
Housing
Mains cable
Accessories
Transducer
The nebulizer chamber should be washed with distilled water after each use. The nebulizer
chamber should be disinfected with normal medical disinfectants such as Buraton 10F as in
the DGHM list.
Note: If non-colour-fast protective wrap is used the plastic parts can become discoloured.
Note: The nebulizer chamber, the nebulizer chamber lid, the continuous filling kit and the
hoses are consumable items. The material they are made of will wear at a greater or
lesser rate depending on the cleaning process used. Before each use, check that all
parts are in good condition. If damage is seen they should be replaced.

Replacement of transducer

Carefully screw on the transducer to the correct tightness. If it is screwed on too loosely,
fluids can escape through the thread; if made too tight, the thread of the plastic chamber
is damaged by cracking due to stress.

Replacement of bacterial filter

- The bacterial filter has to be changed once a week or according to the filter manufacturer.
- If you use a closed sterile water system please change the filter 48 hours after use or every
14 days.
- The bacterial filter has to be changed immediately after contact with water or other
noticeable problems (colour change, breakage, bacteria etc)

Cleaning the single / multiple-use hoses

Multiple-use hoses must be suitably cleaned daily or with change of patient. Single-use
hoses must be replaced daily or with change of patient.
Cleaning
Disinfection
In warm
Immerse or
water with
boil in diluted
mild house-
solution
hold cleaner
Single-use items: Replace after at most 1 week or according
to manufacturer's specific instructions. If a closed sterile
water system is used, filter change is recommended after
48 hours of operation or every 14 days.
Single-use items: Intended for only a single use
Wipe
Wipe
with a
disinfection
damp
(DGHM list)
cloth
Machine
Steam
washing
sterilizing
Wash cycle
Up to 134°C
up to 95°C
5 min
Not
Not
permitted
permitted