SPECIFICATIONS
Performance
Electrical
Environmental
Physical
^
PERFORMANCE
\~j
Range
**
Saturation
\m>
0-100%
\m*
Vij/
K0
Vat*
Pulse Rate
20-250 beats per minute (bpm)
Accuracy
SP021
Adults
70-100%
± 2 digits2
50-69%
± 3 digits2
0-49%
unspecified
Neonates
70-95%
± 3 digits2
Pulse rate
20-250 bpm
± 3 digits2
1.
The reference for accuracy testing is an N-200 oximeter and D-25
sensors that have been validated in human blood studies against
an Instrumentation Laboratories CO-Oximeter. Adult accuracy
specification is based on testing with D-25 sensors; neonatal
accuracy specification is based on testing with N-25 sensors. For
specifications with other NELLCOR sensors, see the sensor
directions for use.
2.
This variation equals one standard deviation (SD). Plus or minus
one SD encompasses 68% of the population.
31