Guide to Operation
Loss of pulse is handled as follows:
The PULSE SEARCH indicator flashes if no pulse is
detected for 3 seconds or for the period during which five
pulses would be expected (whichever is longer). The last
valid measurement continues to be displayed.
The displays go to zero if no pulse is detected for
10 seconds.
The N-20/N-20P turns off if no pulse is detected for
3 minutes.
Changing Operating Mode
In the spot-check mode, hold the measure button down for
3 seconds to change to extended mode.