SECTION 8: SPECIFICATIONS

PERFORMANCE
Range
Accuracy
Performance
Environmental Requirements
Electrical Characteristics
Physical Characteristics
Saturation
0–100%
Pulse Rate
20–250 bpm (beats per minute)
Saturation
±1 standard deviation)
(%SpO
2
SpO
2
Adults
70–100 %
50–69 %
0–49 %
Neonates
70–95 %
Pulse Rate
±3 bpm
20–250 bpm
Note:
The monitor pulse oximeter is calibrated to read oxyhemoglobin sat-
uration (%SpO
dysfunctional hemoglobins (for example) carboxyhemoglobin, methe-
moglobin) may affect the accuracy of the monitor. Indocyanine green,
methylene blue, and other intra-vascular dyes, depending upon their
concentrations, may interfere with the accuracy of the monitor.
Instrument performance may also be compromised by excessive pa-
tient movement, electrosurgical interference, or intense environmen-
tal illumination.
±2 digits
±3 digits
unspecified
±3 digits
) of functional hemoglobin. Significant levels of