Download Print this page

ADC Adtemp ULTRA 417 Use, Care & Maintenance page 4

Digital fever thermometer

Advertisement

1.
Turn the unit OFF, wait 1 minute and then turn ON again.
2.
Retake temperature using proper technique while ensuring normal room temperature (between 73°F -
77°F / 23°C - 25°C).
3.
If unit still reads LO, this indicates that the patient has low body temperature.
WARNING! A low body temperature requires immediate medical attention.
Display: Flickering HI (unit will beep 7 times):
Indication: The measured temperature is above 107.6°F (42°C):
1.
Turn the unit OFF, wait 1 minute and then turn ON again.
2.
Retake temperature using proper technique.
3.
If unit still reads HI, this indicates that the patient has an extremely
high temperature.
WARNING! A high temperature requires immediate medical attention.
Display: Er 1 (unit will beep 5 times and power off automatically):
Indication: The thermometer is too cold or too hot to operate:
1.
Verify room temperature is within product range 60.8°F-104°F (16° C - 40° C).
2.
Wait 1 minute, retake temperature using proper technique.
Display: Er 2 (unit will beep 5 times and power off automatically):
Indication: Temperature did not register because of poor contact, patient movement or poor
placement.
1.
Turn the unit OFF, wait 1 minute, turn it ON again and retake temperature using proper
technique. (See Operating Instructions.)
Display: Er 3 (unit will beep 5 times and power off automatically):
Indication: Unit malfunctioned.
1.
Do not use the thermometer – contact Customer Service. (See last page.)
Display: Flickering battery (after displaying the last temperature measurement)
Indication: Low battery.
1.
The battery has enough remaining life for approximately 10 more measurements. It is
recommended to replace the battery as soon as possible.
Display: Battery icon is on (unit will beep 5 times and power off automatically)
Indication: Drained battery.
1.
The battery is drained - replace battery. (See below on Replacing Batteries.)
Care and Maintenance:
Wipe the tip and probe with an alcohol swab or a cotton swab moistened with 70% isopropyl alcohol
before and after each use. After cleaning, wait at least 2 minutes before taking a measurement for the unit to
readjust to room temperature.
Use only a soft dry cloth to clean the body of the thermometer (if necessary). Never use an abrasive
cleanser or submerge the thermometer in water or other liquids.
Do not expose the thermometer to extreme temperatures, humidity, direct sunlight or shock.
Store at room temperature.