Guidance and manufacturers declaration- electromagnetic Immunity
The PRO 6000 is intended for use in the electromagnetic environment
specified below. The customer or the user of the PRO 6000 should ensure
Immunity
IEC60601
test
test level
Electrostatic
±6kV contact
discharge (ESD)
IEC 61000-4-2
Radiated RF
3 V/m 80MHz to
IEC 61000-4-3
Conducted RF
3Vrms 150kHz to
IEC 61000-4-6
Electrical fast
±2kV power line
transient
± 1kV I/O lines
IEC 61000-4-4
Surge
± 1kV differential
IEC 61000-4-5
± 2kV common
Power frequency
Magnetic field
IEC 61000-4-8
Voltage dips,
short interrupts and
> 95% dip 0.5 cycle
voltage variations
60% dip 5 cycles
on power supply
70% dip 25 cycles
input lines
95% dip 5 sec.
IEC61000-4-11
that it is used in such an environment.
Compliance
level
Complies
±8kV Air
Complies
2.5GHz
Not applicable
(no electrical
80Mhz
cabling)
Not applicable
Not applicable
3 A/m
Complies
Not applicable
Electromagnetic environment
Floors should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be
at least 30%
Field strengths outside the shielded
location from fixed RF transmitters,
as determined by an electromagnetic site
survey, should be less than 3 V/m.
Interference may occur in the vicinity of
equipment marked with the following symbol:
Separation distance calculation provided
below. If a known transmitter is present the
specific distance can be
calculated using the equations.
The PRO 6000 is solely battery powered
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial or
hospital environment
The PRO 6000 is solely battery powered
30