10.3 Manual timer
The Braun PRO 6000 includes a 60 second timer that features an audible notification and visual
indicator at 0, 15, 30, 45, and 60 seconds. The timer turns off automatically 5 seconds after
60 seconds is completed. The timer can be stopped at any time by pressing the Timer button or
applying a probe cover. This feature can be used to time pulse, respirations, etc. To use this feature:
1
Press and hold timer button
A beep is heard at initiation of the timer.
The display will show timer counting up in seconds.
The display will show an icon that has four 15 second quadrants.
The timer will beep at the completion of each 15 second interval
to provide audible notification. Then the current segment will
become solid and the next segment will flash.
At 60 seconds a long beep will sound, all quadrants will appear
solid ending the timer function. The thermometer will exit the
timer mode after an additional 5 seconds.
...
15
0 sec
"Beep"
"Beep"
2
To stop the timer at any time, press the timer button.
11. Settings
11.1 Default temperature scale
To set the default temperature scale:
1
Open battery door by sliding spring
latch to the right using your finger
or a pointed object such as a pen.
While holding the latch in the open
position, grasp the battery door and
remove it. Remove batteries and set
aside. Once batteries are removed,
C/F switch is accessible.
2
Slide switch to C or F using a pen or pointed object.
3
Place batteries back into thermometer. Snap the battery
door back into place and ensure the latch returns to the
original latched position. Celsius or Fahrenheit symbol will
appear on the display.
for one second to activate timer.
...
...
30
45
"Beep"
"Beep"
...
60
"Beep"
or