Ear Temperatures
with the Braun ThermoScan
Ear Canal
98.4
99.1
97.8
94.1
Probe Tip
Tympanic
Membrane
©2002 Welch Allyn, Inc.
All rights reserved
1. Remove the thermometer from the case and
attach a probe cover. The device will turn on.
2. Wait for the "ready" beep and place the
probe snugly into the ear canal; correct probe
placement is essential for accurate readings.
3. Press and release the Start button. The
thermometer will beep once. The green
ExacTemp
light above the Start button will
™
flash, indicating stable probe placement.
4. A long beep and a solid green ExacTemp light
will signal the end of the measurement.
5. If the probe is not placed in a stable position
while taking a temperature, the device beeps,
the green ExacTemp light goes out and POS
(Position Error) will display.
6. Press the
temperature.
7. To change default temperature scale: With
the device off, press and hold the
for about 5 seconds. Release the
when desired scale is shown.
PRO 4000
®
button to recall the last
button
button