Download Print this page
Braun WelchAllyn ThermoScan PRO 4000 Service Documentation

Braun WelchAllyn ThermoScan PRO 4000 Service Documentation

Hide thumbs Also See for WelchAllyn ThermoScan PRO 4000:

Advertisement

Quick Links

6021221_PRO4000_S1 Seite 1 Freitag, 17. März 2006 11:19 11
PRO 4000
Type 6021

Advertisement

loading

Summary of Contents for Braun WelchAllyn ThermoScan PRO 4000

  • Page 1 6021221_PRO4000_S1 Seite 1 Freitag, 17. März 2006 11:19 11 PRO 4000 Type 6021...
  • Page 2 (± 0.5 ° F ). Probe tip: heated up to approximately 37°C. Power supply: rechargeable Braun battery-pack with base-station and smart plug or 2 1.5 V type AA (LR 6) batteries. Smart plug: 100-240 V - input / 12 V – output.
  • Page 3: Service Documentation

    Service Documentation Welch Allyn Part Number: 701627 Rev. G General hints Production code Near the battery storage LOT 4042. 4 = Year / 042 = Day. Attention: Do not dispose of empty batteries or defective rechargeable batteries in household waste. Disposal can take place at appropriate collection sites provided in your country Hint: Type BF equipment This appliance conforms to the following standards: DIN EN 60601-1...
  • Page 4 The Model 9600 PLUS Calibration Tester can be ordered through Welch Allyn Customer Service (see page 8) for Welch Allyn contact information. Model Welch Allyn P/N Braun P/N 9600 PLUS Calibration Tester 01802-110 The following additional data will be necessary:...
  • Page 5 Special care has to be taken, to ensure that only products are handed back to the consumer, which are safe and perform according to its specifications. Furthermore, you are obliged to notify Braun Corporate Quality of any unusual defects of the appliance (see also section "customer complaints").
  • Page 6 Service Documentation Welch Allyn Part Number: 701627 Rev. G Calibration general: Checking the measuring accuracy with blackbody BB 3200 or 9600 PLUS NOTE: It is recommended that the PRO-4000 have an Operational Verification performed annually using the 9600 Plus Blackbody device. •...
  • Page 7 Service Documentation Welch Allyn Part Number: 701627 Rev. G Calibration Mode: To enter the calibration mode, perform the following steps: • Use a new probe cover for calibration check. • Turn the product on by push the -button. Following symbols and functions will be shown on the display.
  • Page 8 Service Documentation Welch Allyn Part Number: 701627 Rev. G Calibration: Follow the operating instructions for the BB-3200 or 9600 PLUS closely. The measured value must be within: 39.7 ° C to 40.1 ° C (103.5 ° F to 104.3 ° F ). NOTE: The measured temperature will vary with ambient temperature.
  • Page 9 Service Documentation Welch Allyn Part Number: 701627 Rev. G Customer complaints: Welch Allyn encourages all service centers to report customer complaints. This information is valuable for the improvements of products. All complaints where it is alleged, that our product has caused or contributed to a death or a serious deterioration in the state of health of a patient, user or other person, must be reported immediately to your local Welch Allyn Organization.
  • Page 10 Service Documentation Welch Allyn Part Number: 701627 Rev. G Authorized Welch Allyn The following is a listing of authorized Welch Allyn Service Centers and Customer Service contact numbers: United States Service Center Welch Allyn North Dock 4619 Jordan Road Skaneateles Falls, NY 13153 Customer Service: 800-535-6663 Service Fax: 315-685-3361...