Filter function

only models:
EMS 300-3
EMS 3000-2
EMS 6K0.1
EMS 12K0.1
(example)
EMS-BA-e-1514
This menu item allows the balance to be set according to
specific ambient conditions and measuring purposes.
 In weighing mode keep the PRINT key pressed until [Unit]
appears.
 Press the MODE button several times until „StAbiL" is
displayed.
 Acknowledge using SET key, the current setting is
displayed.
 Select the desired settings by pressing the MODE button.
Filter 1:
1
The balance reacts quickly and in a sensitive manner, quiet set-
up location.
2
Filter 2:
The balance reacts normally, normal installation site
Filter 3:
3
The balance reacts slowly and in a robust manner, busy set-up
location
 Use the SET key to confirm selection.
27