Adding Supplemental Oxygen
Oxygen can be added to the patient circuit. Please note the warnings listed below when using oxygen with the device.
Warnings:
• When using oxygen with this system, the oxygen supply must comply with local regulations for medical oxygen.
• Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of an open flame.
• When using oxygen with this system, a Philips Respironics Pressure Valve must be placed in-line with the patient
circuit between the device and the oxygen source. The pressure valve helps prevent the backflow of oxygen from
the patient circuit into the device when the unit is off. Failure to use the pressure valve could result in a fire hazard.
Note: Refer to the pressure valve's instructions for complete setup information.
• When using oxygen with this system, turn the device on before turning on the oxygen. Turn the oxygen off
before turning the device off. This will prevent oxygen accumulation in the device.
• Do not connect the device to an unregulated or high pressure oxygen source.
Supplying DC Power to the Device
A Philips Respironics DC power cord can be used to operate this device in a stationary recreational vehicle, boat,
or motor home. In addition, a Philips Respironics DC battery adapter cable, when used with a DC power cord,
allows the device to be operated from a 12 VDC free-standing battery.
Caution: Always ensure that the DC power cord securely fits into your therapy device prior to use. Contact
your home care provider or Philips Respironics to determine if you have the appropriate DC cord for your
specific therapy device.
Caution: When DC power is obtained from a vehicle battery, the device should not be used while the vehicle's
engine is running. Damage to the device may occur.
Caution: Only use a Philips Respironics DC Power Cord and Battery Adapter Cable. Use of any other system
may cause damage to the device.
Refer to the instructions supplied with the DC power cord and adapter cable for information on how to operate
the device using DC power.

Traveling with the System

When traveling, the carrying case is for carry-on luggage only. The carrying case will not protect the system if it is put
through checked baggage. If traveling with the optional humidifier, do not travel with water in the water tank.
For your convenience at security stations, there is a note on the bottom of the device stating that it is medical
equipment and is suitable for airline use. It may be helpful to bring this manual along with you to help security
personnel understand the DreamStation device.
If you are traveling to a country with a line voltage different than the one you are currently using, a different power
cord or an international plug adaptor may be required to make your power cord compatible with the power outlets
of the country to which you are traveling. Contact your home care provider for additional information.
Airline Travel
The device is suitable for use on airlines when the device is operating from an AC or DC power source.
Note: It is not suitable for airline use with any of the modems or humidifiers installed in the unit.