Cleaning The Device; Cleaning Or Replacing The Filters; Cleaning The Tubing; Service - Philips DreamStation CPAP User Manual

Table of Contents

Advertisement

Cleaning the Device

Warning: To avoid electrical shock, always unplug the power cord from the wall outlet before cleaning the device.
DO NOT immerse the device in any fluids.
1. Unplug the device, and wipe the outside of the device with a cloth slightly dampened with water and a mild
detergent. Let the device dry completely before plugging in the power cord.
2. Inspect the device and all circuit parts for damage after cleaning. Replace any damaged parts.

Cleaning or Replacing the Filters

Under normal usage, you should clean the reusable blue pollen filter at least once every two weeks and replace it
with a new one every six months. The light-blue ultra-fine filter is disposable and should be replaced after 30 nights of
use or sooner if it appears dirty. DO NOT clean the ultra-fine filter.
Caution: Dirty inlet filters may cause high operating temperatures that may affect device performance. Regularly
examine the inlet filters as needed for integrity and cleanliness.
This device has an automatic air filter reminder. Every 30 days, the device will display a message reminding you to
check your filters and replace them as directed.
Note: This message is a reminder only. The device does not detect the performance of the filters nor does it
recognize when a filter has been cleaned or replaced.
1. If the device is operating, stop the airflow. Disconnect the device from the power source.
2. Remove the filter(s) from the device. Refer to the "Installing/Replacing the Air Filters" section of this manual.
3. Examine the filter(s) for cleanliness and integrity.
4. To clean the reusable blue pollen filter, first detach the light-blue ultra-fine filter if applicable, and set aside or
dispose of as needed. Then take the reusable filter to the sink, turn it upside down, and run warm tap water
through the white filter media to rinse away any debris. Next, lightly shake the filter to remove as much water as
possible. Allow the filter to air dry completely before reinstalling it. If the reusable blue pollen filter is torn, replace
it. (Only Philips Respironics-supplied filters should be used as replacement filters.)
5. If the light-blue ultra-fine filter is dirty or torn, replace it.
6. Reinstall the filters. Refer to the "Installing/Replacing the Air Filters" section of this manual.
Caution: Never install a wet filter into the device. You must ensure sufficient drying time for the cleaned filter.

Cleaning the Tubing

Clean the flexible tubing before first use and daily. Disconnect the flexible tubing from the device. For the
15 or 22 mm flexible tubing, gently wash the tubing in a solution of warm water and a mild detergent. Rinse
thoroughly. Air dry. Inspect the tubing for damage or wear. Discard and replace if necessary.
Note: Refer to the humidifier manual for the instructions on how to clean the heated tube.

Service

The device does not require routine servicing.
Warning: If you notice any unexplained changes in the performance of this device, if it is making unusual or harsh
sounds, if it has been dropped or mishandled, if water is spilled into the enclosure, or if the enclosure is broken,
disconnect the power cord and discontinue use. Contact your home care provider.