Troubleshooting - Decon wheel e-move User Manual

Hide thumbs Also See for e-move:
Table of Contents

Advertisement

TROUBLESHOOTING

PROBLEM
The chair runs sluggishly
Abnormal vibration
The battery has a short
range?
The lights do not come on
hand after you have charged
the battery
Charging stops abruptly
Charging is never completed
The charger becomes
extremely warm
The charger emits a strange
odour or abnormal sound
Decon Wheel AB
Landerydsvägen
314 22 Hyltebruk
Sweden
CHECK
Is the battery charged?
Has the battery been placed
into the holder correctly?
Have you touched the control
lever by mistake?
Has the system switched off
automatically?
Is there enough air in the
tyres?
Is the wheel loose?
Can an abnormal sound be
heard coming from the
driving machinery?
Is the battery fully charged?
Has a fuse blown?
Has the battery been charged
in a suitable way?
Is the temperature around the
battery and the charger
suitable?
Is the battery fully charged?
Is the temperature suitable?
ACTION
Charge the battery or replace
it.
Place the battery into the
holder.
Press the button again. If it
still runs sluggishly, contact
your supplier.
Switch off the system and
switch it on again.
Pump the tyres.
Put it into place correctly.
Contact your supplier.
Recharge or replace the
battery.
Replace the fuse.
Recharge the battery in a
suitable place.
Try to recharge the battery
when the charger and battery
are at a suitable temperature.
If all of the battery lights do
not come on after recharging,
the capacity has de-creased
due to battery deterioration.
Check the battery's capacity
lights.
Wait until the charger and
battery are at a suitable
temperature.
Contact your supplier.
Contact your supplier.
Tel + 46 (0) 345-408 80
Fax +46 (0) 345-408 95
www.decon.se
info@decon.se