Dos 3 (Neutralizer); Setting The Neutralizer Dosage; Priming The Dosage System - Miele Laboratory Glassware Washer G 7883 Operating And Installation Instructions

Laboratory glassware washer
Table of Contents

Advertisement

Programming functions

DOS 3 (neutralizer)

Setting the neutralizer dosage

Set dosage concentration according to the neutralizer manufacturer's
recommendations.
Step
^ Press the "Drying"
appears in the display.
^ Turn the program selector to "Intensive".
^ Press and hold the "Display"
until the required value appears in the
display, e.g., 0.2.
Dosage concentration setting:

Priming the dosage system

The dosage system does not need to be primed (cleared of air) unless:
– the dosage concentration is being set for the first time,
– the neutralizer container was inadvertently completely emptied .
Step
^ Press the "Drying"
appears in the display.
^ Turn the program selector to "Universal".
^ Press the "Display"
^ Exit the programming mode.
Start "Rinse" program.
36
button until E02
button
button until E02
button.
Display / Result
E02
The programmed dosage
concentration appears in the
display, e.g., 0.10 = 0.10 %
0.20 The dosage concentration is
now set.
Display / Result
E02
d§3
The dosage system will be primed
automatically. Priming is complete
when 0 appears in the display.
This will rinse away any agents that
entered the wash cabinet during
priming.