What if...
Before the start of a
program, the "8
Dosing Pump"
indicator flashes, and
the program can not be
started.
A few minutes after the
start of a program, the
"8 Dosing Pump"
indicator flashes, and
the program stops.
Frequently asked questions
Possible fault
Error message: Fdo:
The liquid detergent
container is empty.
Error message: Fdo:
There is a problem
dispensing the liquid
detergent.
Solution
Before addressing the
error:
– Turn the program
selector to f. The error
message is erased.
– Press the On/Off button
to switch off the
machine.
Refill or replace the liquid
detergent container.
Then:
– Switch the washer on.
– Restart the program.
Before addressing the
error:
– Turn the program
selector to f. The error
message is erased.
– Press the On/Off button
to switch off the
machine.
Refill or replace the liquid
detergent container.
Prime the dosage system
(see "Programming
functions").
Then:
– Turn the washer on.
– Restart the program.