Plumbing
– Connect the intake hose for the
steam condenser (without water
safeguard) to the cold water faucet.
– See the next page for how to
connect the DI water line.
– Water valves with
to be provided on site. They should
be easily accessible, because the
water supply will need to be turned
off when not in use.
– The inlet hoses are approx. 5 ½ ft
(1.7 m) long pressure hoses with
3
/
inch thread. Do not remove the
4
water inlet filters in the threads.
– Install the large surface area filters
(in the kit supplied with the machine)
between the water valve and the
water inlet hose (see illustration in
"Cleaning the water inlet filters"). The
large surface area filter for distilled
water is made of stainless steel and
can be recognized by the matte
surface.
66
3
/
inch thread are
4
Do not cut or damage the inlet hose
in any way.
Please also refer to the installation
diagram.