AC voltage verification test

Table 1-7 AC volts verification test

Vrms
Input Frequency
100 mV
20 Hz
100 mV
45 Hz
100 mV
1 kHz
100 mV
10 kHz
100 mV
30 kHz
100 mV
100 kHz
1 V
20 kHz
1 V
45 kHz
34450A Service Guide
Configuration: AC Volts (CONFigure[:VOLTage]:AC)
1 Connect the calibrator to the front panel HI and LO input terminals
as shown in the figure below:
2 Select the AC voltage function.
3 Select each range in the order shown below. Provide the indicated
input voltage and frequency. Compare measurement results to the
appropriate test limits shown in the table. (Be certain to allow for
appropriate source settling.)
Function
AC voltage
Performance Tests and Calibration
[1]
Range
100 mV
100 mV
100 mV
100 mV
100 mV
100 mV
1 V
1 V
Performance Verification Tests
Error from
Quick Check
Nominal 1 Year
1
±1.1 mV
±0.3 mV
±0.3 mV
±0.3 mV
±1.8 mV
±3.3 mV
±11 mV
±3 mV
17