Monitor Reassembly; Figure 6-3: Separating Case Halves - Nellcor NPB-190 Service Manual

Pulse oximeter
Table of Contents

Advertisement

Section 6: Disassembly Guide
6.5

MONITOR REASSEMBLY

6-4

Figure 6-3: Separating Case Halves

1.
Place the two halves of the monitor on the table as shown in Figure 6-3 and
connect the Power Supply to J6 on the UIF PCB.
2.
Place the top case over the bottom case and align the four outside screw
posts and close the monitor.
Caution: When reassembling the NPB-190, hand tighten the screws that
hold the cases together to a maximum of 10 inch-pounds. Over-tightening
could strip out the screw holes in the top case, rendering them unusable.
3.
Install the four corner screws.
J6
Power supply