Figure 3-1: Npb-190 Controls - Nellcor NPB-190 Service Manual

Pulse oximeter
Table of Contents

Advertisement

Section 3: Performance Verification
3.3.2 Performance Tests
3.3.2.1
Power-On Self-Test
3-2
2.
Verify that the monitor is off and that the AC Power/Battery Charging
indicator is lit.
3.
Charge the battery for at least 14 hours.
The power-up performance tests (3.3.2.1 and 3.3.2.2) verify the following
monitor functions:
Power-On Self-Test
Factory Power-On Defaults and Alarm Limit Ranges
On/Standby
Lower Alarm
Limit
Note:
Refer to Figure 3-1, NPB-190 Controls, when following the instructions
listed below.
1.
Connect the monitor to an AC power source. Verify that the AC
Power/Battery Charging indicator is lit.
2.
Do not connect any input cables to the monitor.
3.
Observe the monitor front panel. With the monitor off, press the Power
On/Standby button. Verify that the monitor performs the following
sequence:
Upper Alarm
Limit

Figure 3-1: NPB-190 Controls

Alarm Silence
NPB-190
Adjust
Adjust
Down