Product Specifications - Braun IRT 3520 User Manual

Hide thumbs Also See for IRT 3520:
Table of Contents

Advertisement

Available languages

Available languages

Product specifications

Displayed temperature range
Operating ambient
temperature range
Display resolution
Accuracy for patient temperature range
96.8 °F – 102.2 °F (36 °C – 39 °C): ± 0.4 °F
outside this range:
Long term storage ranges
Temperature:
Humidity:
Battery life:
This infrared thermometer meets requirements established in
ASTM Standard E 1965-98. Full responsibility for the conformance
of the product to the standard is assumed by
Braun GmbH, 61476 Kronberg, Germany.
ASTM laboratory accuracy requirements in the display range of
98.8 °F to 102.2 °F (36 °C to 39 °C) for infrared thermometers is
± 0.4 °F (± 0.2 °C), whereas for mercury-in-glass and electronic
thermometers, the requirement per ASTM Standards E 667-86
and E 1112-86 is ± 0.2 °F (± 0.1 °C).
A clinical summary is available upon request.
The Braun ThermoScan thermometer has been proven to be safe.
To ensure the best results, please take the time to read this manual
completely and keep it handy for future reference.
In the unlikely event you experience any difficulty using the Braun
ThermoScan thermometer, simply call us toll free at:
1 (800) 327-7226.
TYPE BF
Attention, consult
EQUIPMENT
ACCOMPANYING CLASSIFIED BY UNDERWRITERS LABORATORIES INC.
DOCUMENTS
Internally Powered Equipment
Continuous Operation
Not Protected against Ingress of Water
U.S. Patent No. 5,088,834 Other Patents Pending
93.2 °F – 108 °F (34 °C – 42.2 °C)
50 °F – 104 °F
0.1 °F or °C
Maximum Laboratory Error
± 0.5 °F
– 4 °F to 122 °F ( – 20 °C to 50 °C)
95 % non-condensing
2 years / 1000 measurements.
MEDICAL ELECTRICAL EQUIPMENT
WITH RESPECT TO ELECTRIC SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH UL 2601-1 / CAN CSA C22.2
No.601.1 10HK
(10 °C – 40 °C)
± 0.2 °C
± 0.3 °C
®
15

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents