The Braun ThermoScan® thermometer is indicated for intermittent measurement and monitoring of human body temperature for people of all ages.
Please consult your doctor if you see symptoms such as unexplained irritability, vomiting, diarrhea, dehydration, changes in appetite or activity, seizure, muscle pain, shivering, stiff neck, pain when urinating, etc. in spite of absence of fever.
The operating ambient temperature range for this thermometer is 10 – 40°C (50 – 104°F).
Do not expose the thermometer to temperature extremes (below –25°C / –13°F or over 55°C / 131°F) or excessive humidity (> 95% RH).
This thermometer must only be used with genuine Braun ThermoScan® disposable lens filters. Other lens filters can lead to inaccuracy.
To avoid inaccurate measurements always use this thermometer with a new, clean, disposable lens filter.
If the thermometer is accidentally used without a disposable lens filter, clean the lens after use (see «Care and cleaning»).
Keep disposable lens filters out of reach of children.
This thermometer is intended for household use only.
Use of this thermometer is not intended as a substitute for consultation with your physician. This thermometer must stay in stable ambient (room) temperature for 30 minutes before operating.
Thermometer will signal by beeps normal, elevated or high temperature.
Temperature Reading | Number of Beeps |
Normal (≤37.4°C or 99.3°F) | One beep: ) |
Elevated (≥ 37.5°C or 99.5°F) | Two beeps, repeated: )) )) )) |
High (≥ 38.6°C or 101.5°F) | Four beeps, repeated: )))) )))) )))) |
Normal body temperature is a range. It varies by site of measurement and it tends to decrease with age. It also varies from person to person and fluctuates throughout the day. Therefore, it is important to determine normal temperature ranges.
Before taking the temperature, make sure the ear is free from obstructions or excess earwax.
Do not attempt to take the reading without a disposable lens filter on – the measuring sensor must be kept protected and clean from earwax to give accurate readings.
Change the disposable lens filter with each use to ensure hygiene, to keep child safe from cross-contamination and to ensure earwax residues on the disposable lens filter do not impact accuracy of the reading.
Use a clean disposable hygienic lens filter with each measurement:
Earwax and dirt on the tip of the disposable lens filter, even in microscopic layer, can create an obstacle between the sensor and the infrared heat emitted by the ear and impact the accuracy of the measurement.
The probe tip (5) is the most delicate part of the thermometer. It has to be clean and intact to ensure accurate readings.
If the thermometer is accidentally used without a disposable lens filter, clean the probe tip as follows:
Very gently wipe the surface with a cotton swab or soft cloth moistened with alcohol. Do not use cleaners other than alcohol on the probe tip. Use of other cleaners could damange the tip.
Once the alcohol has dried completely, a new disposable lens filter can be put on and a temperature measurement taken. If the probe tip is damaged, please contact your authorized local service center.
Use a soft, dry cloth to clean the thermometer display and product exterior. Do not use abrasive cleaners. Never submerge this thermometer in water or any other liquid.
Store the thermometer and disposable lens filters in a dry location free from dust and contamination and away from direct sunlight.
Additional disposable lens filters (LF 40) are available at most stores selling Braun ThermoScan.
Situation | Solution |
No disposable lens filter is attached. | Attach new, clean, disposable lens filter. |
Measurement before device stabilization | Wait until all the icons stop flashing. |
Ambient temperature is not within the allowed operating range (10–40°C or 50–104°F). | Allow the thermometer to remain for 30 minutes in a room where the temperature is between 10 and 40°C or 50 and 104°F. |
HI = too high LO = too low Temperature taken is not within typical human temperature range (34–42.2°C or 93.2–108°F) | Make sure the probe tip and lens are clean and a new, clean, disposable lens filter is attached. Make sure the thermometer is properly inserted. Then, take a new temperature. |
System error | Reset the thermometer by removing the battery and then replacing it again, wait 1 minute, then turn on again. Please contact your authorized Service Center (See Guarantee Card). |
Battery is low. | Insert new battery. |
Battery is too low. Device cannot be powered on to the ready state. | Insert new battery. |
Do you have any further questions? | Please contact your authorized Service Center (See Guarantee Card). |
Make sure the thermometer is turned off. Press and hold down the <<start>> button. Keep holding down the <<start>> button and after about 3 seconds the display will start to cycle between temperature scales (°C or °F).
Release the start button when the desired scale is shown. This will reset the thermometer to the new temperature scale.
The thermometer is supplied with one 3V Lithium cell battery (type 2032). Insert new battery when the low battery symbol appears on the display.
The battery door design is intended to prevent a small child from easily opening the battery door to address a possible choke hazard. Open the battery compartment by inserting a narrow pointed object such as a ball point pen or paperclip into the small hole above the battery door (8). While pressing into the hole slowly slide back the battery door with the other hand.
To close battery door align the tabs along the side of the battery into the battery compartment and slowly slide back into position until the door clicks into its locked position.
This product has been calibrated at the time of manufacture. If used according to the use instructions, periodic calibration is not required. If at any time, you question the measuring accuracy, please contact our service representative (see last page for contact information) or visit www.hot-europe.com/support.
Manufacturing date is given by the LOT number located in the battery compartment.
The first three (3) digits after LOT represent the Julian date that the product was manufactured and the next two (2) digits represent the last two numbers of the calendar year the product was manufactured. The last identifiers are the letters that represent the manufacturer.
An example: LOT 11614fam, this product has been manufactured on the 116th day of the year 2014.
EMC tables and Consumer Card are available on our website at www.hot-europe.com/support
For support requests, please visit www.hot-europe.com/support or find service contact information on the last page of this manual.
UK only: This guarantee in no way affects your rights under statutory law.
Model: | IRT 3030 |
Displayed temperature range: | 34°C – 42.2°C (93.2°F – 108°F) |
Operating ambient temperature range: | 10°C – 40°C (50°F – 104°F) |
Storage temperature: | -25°C – 55°C |
Display resolution: | 0.1°C or °F |
Accuracy for displayed temperature range: | ± 0.2°C (35.5 – 42°C) (95.9 – 107.6°F) ± 0.3°C (outside this temperature range) |
Clinical bias: | -0.11°C (-0.198°F) |
Limits of agreement: | ± 0.25°C (0.45°F) |
Clinical repeatability: | ± 0.10°C (0.18°F) |
Measuring Site: | Ear |
Reference Site: | Oral (estimated oral temperature is displayed) |
Operating and storage relative humidity: | ≤ 95% non-condensing |
Battery life: | 2 years /1000 measurements |
Service life: | 5 years |
Warranty: | 2 years |
If device is not used or stored within specified temperature and humidity ranges the technical accuracy cannot be ensured.
Type BF equipment
See instructions for use
Operating temperature
Storage temperature
Subject to change without notice.
IP22: Protected against solid foreign objects of 12.5 mm diameter and greater. Protected against vertically falling water drops when the device is tilted up to 15 degrees.
EN 12470-5:2003 Clinical thermometers—Part 5: Performance of infra-red ear thermometers (with maximum device)
ISO 80601-2-56:2009 Medical electrical equipment— Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
This product conforms to the provisions of the EC directives 93/42/EEC (Medical Device Directive) and 2011/65/EU (RoHS).
The Braun ThermoScan needs special precautions regarding EMC. For detailed description of EMC requirements please refer to www.hot-europe.com/support.
Portable and mobile RF communications equipment can affect the Braun ThermoScan.
This product contains batteries and recyclable electronic waste. To protect the environment, do not dispose of it in the household waste, but take it to appropriate local collection points.
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Advertisement
Thank you! Your question has been received!
Need Assistance?
Do you have a question about the ThermoScan IRT 3030 that isn't answered in the manual? Leave your question here.