Product Specifications - Braun ThermoScan PRO 3000 Manual

Hide thumbs Also See for ThermoScan PRO 3000:
Table of Contents

Advertisement

Available languages

Available languages

14
Product specifications
Displayed temperature range
68 °F - 108°F (20°C - 42.2°C)
Operating ambient
temperature range
50 °F - 104 °F (10°C - 40°C)
Display resolution
0.1°F or °C
Accuracy for patient temperature range
Maximum Laboratory Error
96.8 °F - 102.2 °F (36°C - 39°C):
+0.4°F
+0.2°C
outside this range:
+0.5°F
+0.3°C
Long term storage ranges
Temperature:
~4°F to 122°F (-20°C to 50°C)
Humidity:
95% non-condensing
Shock: Withstands one meter drop
Measurement time
1 second
Minimum time between readings
2 seconds
Automatic «power down»
30 seconds
Battery life
5,000 temperatures
Battery type
2 x CR 2032/DL 2032
Thermometer dimensions
6.4" x 1.7" x 1.2"
(16.3 cm x 4.4 cm x 3.1 cm)
Thermometer weight
3.5 ounces (100 g)
{without batteries)
This infrared thermometer meets requirements established
in ASTM Standard E 1965-98. Full responsibility for the conform-
ance of the product to the standard is assumed by Braun GmbH,
61476 Kronberg, Germany.
ASTM laboratory accuracy requirements in the display range of
96.8 °F to 102.2 °F (36 °C to 39°C) for infrared thermometers is
+0.4 °F (+0.2°C), whereas for mercury-in-glass and electronic
thermometers, the requirement per ASTM
Standards E667-86
and E1112-86 is +0.2 °F (+0.1°C).
A clinical summary is available upon request.

Advertisement

Table of Contents
loading

Table of Contents