SERVICE MANUAL
8.9.1.1

Replacing the tilt module microswitch

Requirements:
● socket spanner, 10 mm
● 5 mm Allen key
● oblique pliers
● cable ties
Note
Either two microswitches, or one microswitch and an angular sensor, can be fitted. Replacing
the angular sensor is described in Chapter 8.9.1.2.
Removing the tilt module microswitch:
Undo and remove the two cap nuts (1) on the right-hand plastic cover (2) with a 10 mm socket
spanner.
Remove the plastic cover.
Remove the two washers (5).
Loosen and remove the two nuts (4) with a 10 mm socket spanner.