Loosen and remove the front bolt (1) using
the 6 mm Allen key.
Remove locking-gas cylinder.
Dispose old locking-gas cylinder (1).
Install new locking-gas cylinder (2) in
wheelchair.
Locking-gas cylinder rear mount.
Provide 2 screws (1) with thread-locking
fluid.
Tighten the 2 screws (1) with 10 Nm
Push in and fix gas cylinder cam (1) e.g.
using a screw driver (2).
Provide screw (5) with thread-locking fluid.
IMPORTANT:
Install brass bush (6) and washer (4) and
screw (5) in a way that the bush (6) is on the
walking beam side - thus on the outside of
the wheelchair.