Fault Diagnosis With Electric Actuator Motors - Invacare Typhoon II Service Manual

Invacare typhoon ii wheelchair service manual
Hide thumbs Also See for Typhoon II:
Table of Contents

Advertisement

6.1.2

Fault diagnosis with electric actuator motors

Please use the following table to assess fault causes when using electric actuator motors.
PROBLEM
Electric
actuator
motor does
not react
18
OTHER
POSSIBLE
SYMPTOMS
CAUSE
Remote shows
Lighting /
blinking "E",
actuator
status diode on
module
lighting/actuator
defective
module does not
go out even if
the remote is
switched off or
disconnected.
None
Cable
possibly
disconnected
or damaged
Electrical
actuator
motor
possibly
defective
Remote
possibly
defective
SOLUTION
Documentation
• Replace
See "Replacing
lighting/actuator
electronic
module
components" on
page 31
• Check to
See "Checking the
cables" on page 41
ensure that the
cable has not
been
disconnected
or damaged.
Replace cable
if necessary
• Test actuator
See "Checking an
motor
actuator motor" on
page 45
• Replace the
See "Replacing the
remote on the
ACS Remote" on
wheelchair in
page 42
order to rule
out the
possibility that
the remote is
causing the
fault.