Invacare Typhoon II Service Manual page 37

Invacare typhoon ii wheelchair service manual
Hide thumbs Also See for Typhoon II:
Table of Contents

Advertisement

CAUTION! Fire hazard!
Cables can get jammed and frayed.
• Ensure the cables have the correct polarity! They must not protrude into the lifter area. Use
cable clamps if necessary.
When installing new batteries, cables must never be routed between the front battery and the lifter
actuator! If so, they can be damaged when the lifter is operated.
WRONG!
Installation takes place in reverse order.
Please note
The battery terminals on the rear battery must face the rear and terminals on the front battery
must face the front. The batteries cannot be connected in any other fixing direction.
WARNING: Danger of crushing!
Danger of injury to your hands!
• Ensure that the actuator head slides into the top holder.
Run the actuator bolt into the top position
and ensure that it slides into the top holder
(1).
RIGHT!
37