Setting Initialization

Preset measurement conditions can be initialized to the factory default settings according
to the following procedure.
Factory default settings are shown below.
* Synchronizing method
* Observer
* Color space mode
* Lens type
* Backlight during measurement : ON
Operation Procedure
1
. When the power switch is OFF (O),
turn it to ON ( I ) while pressing
MEMORY key,
simultaneously.
The INITIALIZE SETTINGS (confirmation of
setting initialization) screen appears approx.
5 seconds after the initial screen on the LCD.
Continue pressing MEMORY key,
key until the INITIALIZE SETTINGS screen
appears.
2
. Press
key to select [OK] and then
press ENTER key.
The preset measurement conditions are
initialized and the MEAS screen appears on the
LCD.
: NO SYNC
* Measurement time
: 2°OBS
* Display format
: L
xy
* Absolute value (ABS)/
v
Difference (DIFF) display
: STANDARD
* ND filter type
1
key and
key
key and
85
: MULTI INTEG, s,
IN-ND: AUTO
: ****.**** [F]
: Absolute value (ABS)
: NONE
1, 2
I N I T I A L I Z E
S E T T I N G S
RESET
ALL SETTINGS ?
OK
CANCEL
I N I T I A L I Z E
S E T T I N G S
RESET
ALL SETTINGS ?
OK
CANCEL