When Using ND Filter

Use the ND filter in the optional accessories when measuring a high-luminance object.
If the ND filter is attached, the measurement value must be compensated for the filter
transmittance. This compensation coefficient is attached to the ND filter. Use the data management
software CS-S0w Professional supplied with the CS-2000 as standard to set the coefficient in this
instrument. After that, when the ND filter is selected, the measurement value compensated with the
compensation coefficient can be obtained. See instruction manual of CS-S0w for details.
The wrong setting of ND filter leads to inaccurate measurement.
Do not use the ND filter with the closeup lens. It leads to inaccurate measurement.
Note that an extra ND filter is built into this instrument. There are three setting options for use or
non-use of this built-in ND filter: [AUTO] for switching automatically depending on the luminance of
the object, [OFF] for constant non-use, and [ON] for constant use (See p.3 to p.33).
* ND-FILTER
: NONE, ND0, ND00
* Factory default setting : NONE
Operation Procedure
1
1
. Press MENU key when the MEAS
screen is displayed.
The MENU screen appears.
When the backlight of the LCD has been turned
off via BACKLIGHT key on the MEAS screen,
the backlight is turned on.
2
. Press either
[OPTION] and then press ENTER
key.
The MENU – OPTION screen appears.
The current set contents are displayed in the
[ND-FILTER] item.
6, 7
or
key to select
45
2, 3, 4, 5
M E A S
S N G L
U C 0 0
Lv
35. 93
x
0 . 3968
y
0 . 4060
Obs
SPD
SYNC
L E N
ND
Nr m
I n t
S t d
N o n
M E N U
MEAS
MEMORY
T A R G E T
O P T I O N
S E T U P
M E N U
O P T I O N
U S E R . C A L
NON
L E N S
STD
ND-F I L T E R
NON
T 0 1
c d
m
2
BL
o n