Error Check - Konica Minolta CS-2000 Instruction Manual

Spectroradiometer
Hide thumbs Also See for CS-2000:
Table of Contents

Advertisement

Error Check

Should any errors be found in this instrument, try the corrective actions shown in the
following table. If this does not help, it's possible the instrument is broken. Please
contact the nearest KONICA MINOLTA authorized service facility with the error
number and version of your instrument. To confirm the instrument version, see p.23.
Error
Symptom
No.

No display on the LCD
after power is turned
on.
2
Nothing is visible
through the finder.
3
Nothing is displayed on
the LCD.
4
Does not accept key
operation.
5
Measurement unavailable
even if the measurement
button is pressed.
6
The entered value for
target color is different
from the one that will be
displayed after setting.
7
Measurement values
do not appear.
Item to Check
Has the AC adapter been properly
plugged into the AC outlet?
Has the AC adapter been
connected to this instrument?
Has a wrong AC adapter
been connected?
Is AC power source
within rated scale?
Is the lens cap still attached
to the objective lens?
Is the ND filter attached
to the objective lens?
Is the ND eyepiece filter
attached to the finder?
Is the backlight set to
OFF?
Has the backlight been
set to OFF during
measurement?
Has the remote mode
been set?
Have you pressed a
disabled key?
Has a screen other than
MENU been displayed?
Is there data?
Has the color space
mode become color
temperature?
Did you interrupt
measurement?
82
Corrective Action
Connect the AC adapter.
Connect the AC adapter.
Be sure to use the AC adapter and
power cord supplied as a standard
or optional accessory (AC-A32).
Use within ±0% of the nominal
voltage.
Remove the lens cap.
Use the ND filter when the
luminance of that being
measured is too high.
Use the ND eyepiece filter when
luminance of that being
measured is too high.
Press BACKLIGHT key to turn
ON the backlight.
In the menu operation, set the
backlight to ON during
measurement
Press ESC key to cancel the
remote mode.
Press the correct key.
Conduct measurement when the
MEAS screen is displayed.
A -digit difference may be
found due to calculation error.
Conduct measurement.
Color temperature is displayed
as "_ _ _" if it is far from the
blackbody locus. Change the
display using the other color
space mode and check.
Conduct measurement again.
Reference
Page
2
2
2
2
8
9, 45
9, 45
4, 49
49
73
53
6
52
39
52