Error Message

Error message appears on the LCD when this instrument does not operate normally. Below
table shows type of error message, its description and corrective action respectively.
Error Message

MEMORY ERROR
2
OVER
3
SYNC ERROR
4
VIEWING-ANGLE
SELECTOR ERROR
5
TEMPERATURE
ERROR
6
FAN ERROR
Cause (Description)
Data stored in ROM is
damaged.
Luminance of measuring
object is higher than the
available range.
Flicker of measuring object
is high.
Input signal in EXT SYNC
mode is not at CMOS level.
Input signal in EXT SYNC
mode exceeds 200 Hz.
Input signal in EXT SYNC
mode is less than 20 Hz.
Measurement was
performed when the
measuring angle selector
was in the wrong position,
or its position was changed
during measurement.
The ambient temperature for the
measurement device is too high,
and an internal temperature of
the sensor becomes abnormal.
The cooling fan stops, or the
cooling device has an error.
8
Corrective action
Turn on the power switch again.
If the symptoms are not remedied,
please contact the nearest
KONICA MINOLTA
authorized service facility.
• Use the ND filter and re-measure.
• Decrease the measuring area and
re-measure.
• If the symptoms are not remedied,
please contact the nearest
KONICA MINOLTA authorized
service facility.
• Set the flicker cycle in INT SYNC
mode, or input the flicker signal in
EXT SYNC mode.
• Input a signal with 3.5 to 5.0 V at
high level; 0.0 to .5 V at low level.
• Set the frequency value divided by
an integer in INT SYNC mode and
re-measure.
• Set the integral multiple value of
the input signal cycle in INT SYNC
mode and re-measure using the
MULTI INTEG-NORMAL mode or
MULTI INTEG-FAST mode.
• Set the integral multiple value of the input
signal cycle in INT SYNC mode and re-
measure using the MANUAL mode.
• Switch the measuring angle
selector and re-measure. Or, do
not operate the measuring angle
selector during measurement.
• If the symptoms are not remedied,
please contact the nearest
KONICA MINOLTA authorized
service facility.
• Decrease the ambient temperature
to enable cooling to within the
specified temperature.
• Decrease the ambient temperature
to enable cooling to within the
specified temperature.
• Turn off the power switch once,
turn it on again after a while.
• If the symptoms are not remedied, the
fan must be repaired. Please contact
the nearest KONICA MINOLTA
authorized service facility.