Chapter 4
Examining Chromatogram Data
Type of Data
Higher resolution
data
Higher molecular
weight data
4
4-18
Applied Biosystems
2. Select the method to use based upon the type of data you
are examining, then enter the associated value displayed
for the method you select:
Suggested
Method
Noise Filter
Specify a Correlation Factor of 0 to 1.0.
(NF)
Settings from 0.5 to 0.7 yield acceptable
results for most data. A setting close to 1.0
yields a higher degree of noise reduction.
(May affect
peak
If applying the Noise Filter with a certain
resolution.)
Correlation Factor does not yield the
necessary noise removal, return to the original
trace (see "Returning to the original trace" on
page 4-4) and apply the Noise Filter again with
a higher Correlation Factor setting. Applying
the filter more than one time with the same
Correlation Factor setting does not improve
noise removal.
Gaussian
Specify a Filter Width in data points (odd
Smooth
integers only). The maximum number of
(SM)
smoothing points is 2001. Points less than 1
Filter Width from the edge of the spectrum are
not smoothed.
(May affect
peak
resolution.)
Description