Applied Biosystems Data Explorer 4 Series User Manual page 124

Data explorer user guide v4 series sw user guide
Table of Contents

Advertisement

Chapter 3
Peak Detection and Labeling
Peak Processing
parameters
(spectrum data
Table 3-3 Peak Processing Parameters (Spectrum Data Only)
Parameter
Integration Baseline Settings
NOTE: You can set peak labels to display peak start, peak end, and baseline
marks. See Section 3.5.2, Setting Chromatogram and Spectrum Peak Labels.
Valley-to-Baseline
3
Valley-to-Valley
Spectrum Parameters
Centroid %
Enable Gaussian
Fitting
3-26
Applied Biosystems
Table 3-3 describes the parameters in the Peak Processing
tab of the Spectrum Peak Detection Setup dialog box (see
Figure 3-5 on page 3-16).
only)
Default peak detection values are listed in Section 3.7, Default
Peak Detection Settings.
Draws a vertical line from all valleys to a horizontal baseline.
The level of the horizontal baseline is determined using the
minimum peak valley point (left or right) for each peak. See
Figure 3-25 on page 3-70.
Forces a baseline through all valley points. See Figure 3-25
on page 3-70.
Specifies the percentage of peak height used to determine
the location of the centroid mass. For example, with a
Centroid % of 10, the software examines the top ten percent
of signal.
Increases the accuracy of the centroid calculation for
spectral peaks with limited data points (fewer than 6 data
points across the peak), in particular, Mariner data below
m/z 400.
Description
Continued