Agilent Technologies XX3625 Service Manual page 49

Table of Contents

Advertisement

±
Equipment Required
Equipment
Critical Specifications
0.5 V to 2 Vdc, ±1 mV accuracy
Power Supply
Digital Multimeter
Better than 0.1% accuracy
(DVM)
Cables (2)
BNC
Adapters (2)
BNC (f) to banana (m)
Adapters (2)
BNC tee (m)(f)(f)
Blocking capacitor
0.18
Shorting cap
BNC
Volts/div
Position
200 mV
2.00000 V
100 mV
1.00000 V
50 mV
500.000 mV
µ
F
Supply
Tolerance
± 36 mV
2.00 V
± 18 mV
1.00 V
± 9 mV
500 mV
Chapter 3: Testing Performance
To test offset accuracy
Recommended Model/Part
Agilent 6114A
Agilent 34401A
Agilent 10503A
Agilent 1251-2277
Agilent 1250-0781
Agilent 10240B
Agilent 1250-0774
Limits
minimum
maximum
1.964 V
2.036 V
0.982 V
1.018 V
491 mV
509 mV

Advertisement

Table of Contents
loading

This manual is also suitable for:

Xx410654835a54845a54846a