Chapter 4: Calibrating and Adjusting
To run the self calibration
Equipment Required
Equipment
Critical Specifications
BNC 50 Ω 9 inch
Cable
Let the Oscilloscope Warm Up Before Running the Self Calibration
The self calibration should only be done after the oscilloscope has run for one half hour at ambient
temperature with the cover installed. Calibration of an oscilloscope that has not warmed up may result in
performance test failure.
Calibration time
It will take approximately 45 minutes to run the self calibration on the oscilloscope, including the time
required to change cables from channel to channel.
± °
Recommended Model/Part
Agilent 10502A